Pipette | Volume | Accuracy | Precision | ||
---|---|---|---|---|---|
%D | µl | %CV | µl | ||
p20 GEN2 | 1 | ±15% | 0.15µL | ±5% | 0.05µL |
10 | ±2% | 0.2µL | ±1% | 0.1µL | |
20 | ±1.5% | 0.3µL | ±0.8% | 0.16µL | |
p300 GEN2 | 20 | ±4% | 0.08µL | ±2.5% | 0.5µL |
150 | ±1% | 1.5µL | ±0.4% | 0.6µL | |
300 | ±0.6% | 1.8µL | ±0.3% | 0.9µL | |
p1000 GEN2 | 100 | ±2% | 2.0µL | ±1% | 1µL |
500 | ±1% | 5.0µL | ±0.2% | 1µL | |
1000 | ±0.7% | 7.0µL | ±0.15% | 1.5µL |
Pipette | Volume | Accuracy | Precision | ||
---|---|---|---|---|---|
%D | µl | %CV | µl | ||
P10 8-channel | 10 | ±2 | ±0.2 | ±3 | ±0.3 |
5 | ±5 | ±0.25 | ±5 | ±0.25 | |
1 | ±5 | ±0.05 | ±10 | ±0.1 | |
P300 8-channel | 300 | ±0.5 | ±1.5 | ±1.5 | ±4.5 |
150 | ±0.8 | ±1.2 | ±2.5 | ±3.75 | |
30 | ±2.5 | ±0.75 | ±5 | ±1.5 |